On january 6, 2021 haemonetics was notified of an error code 236, 3 to 1 manifold tubing not being installed which was observed during an urgent gynecological procedure, utilizing the cell saver(r) elite system cell saver(r) elite set - 125ml in (b)(6).Blood waste volume was around 1800ml (combination of blood and ac solution).It was then necessary to transfuse allogeneic blood to the patient, patient had a reaction to the transfusion.Following the incident, patient was monitored in icu.Follow up indicated that patients condition stabilized and patient was moved out from icu.
|