• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ULTRAFLEX DUO HIGH FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ULTRAFLEX DUO HIGH FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Model Number N/A
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 11/23/2020
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined that the unit had a cracked top plate; however, the repair was not completed because this was a non field repairable issue.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item.However, as the device is only assigned a unique serial number, an additional review could not be performed.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
 
Event Description
It was reported that the cart was leaking during cleaning.The evaluation of this device and follow-up received has clarified that this device was leaking from the top with the manifolds out.No adverse events were reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX DUO HIGH FLUID CART WITH SMOKE EVACUATOR
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key11259831
MDR Text Key229837858
Report Number0001526350-2021-00112
Device Sequence Number1
Product Code JCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133786
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberULDU500SE
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/23/2020
Initial Date FDA Received02/01/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/04/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-