Catalog Number 0603870C
Device Problems Fracture (1260); Suction Problem (2170); Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/16/2020
Event Type  malfunction  
Manufacturer Narrative
Manufacturing review: the device history record review was performed for the reported lot number and this lot meets all release criteria.The manufacturing review did not indicate any possible manufacturing issue that could be related to the reported event.Investigation summary: one mri hard base port attached to a groshong catheter was returned for evaluation and one electronic photo was provided for review.Visual, microscopic visual and functional evaluation were performed.The investigation is confirmed for the reported aspiration issue and catheter fracture and scratched material as the port body and attached catheter were patent to infusion and aspiration with a leak noted at the splits and air was aspirated into the syringe upon aspirating the water, multiple splits were noted approximately 2mm from the distal end of the cath-lock and scoring was noted on the cath-lock.Based upon the available information a definitive root cause could not be determined.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.The catalog number identified has not been cleared in the us but is similar to the m.R.I.Low-profile implantable port, groshong single-lumen, 7f products that are cleared in the us.The pro code and 510 k number for the m.R.I.Low-profile implantable port, groshong single-lumen, 7f products are identified in procode and pma/510k.(expiry date: 11/2022).
 
Event Description
It was reported that one day post port placement procedure through right jugular vein, the device was allegedly unable to be aspirated.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GROSHONG PORT SINGLE LUMEN
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key11264751
MDR Text Key230142164
Report Number3006260740-2021-00178
Device Sequence Number1
Product Code LJT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K924250
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0603870C
Device Lot NumberREDZ0872
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/07/2021
Initial Date FDA Received02/02/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/21/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age41 YR
Patient Weight52
-
-