Model Number CR2
Device Problems Nonstandard Device (1420); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/24/2020
Event Type  malfunction  
Manufacturer Narrative
The customer received a replacement device.The customer's device was archived by physio-control.An engineering investigation has determined that due to the design of the lcpr2 lid switch, absence of the lid magnet allows current to flow from the battery, even when the device is in standby mode.This will reduce the life of the battery.Therefore, the reported issue, the loss of lid/lid magnet, is associated with two hazardous situations: opening the lid does not turn on the device, and, higher than intended current draw while the device is in standby mode results in a prematurely depleted the battery.Replacement of the device lid so that the magnet is present in the device allows the lid switch to function as designed; turning on/off the device when the lid is opened and preventing current draw when the device lid is closed.The lpcr2 operating instructions has been updated to include additional troubleshooting tips to direct the customer to act when device behavior indicates the lid magnet may be missing.A new warning informs the customer of the risk associated with a missing magnet.Device was archived by manufacturer and not evaluated.
 
Event Description
The customer contacted physio-control to report that their device was lid magnet fell out from the lid of the device.In this state the device will not turn on automatically when the lid is activated, which can cause a delay in defibrillation therapy if needed or may lead to use error resulting in a failure to deliver defibrillation.There was no patient use reported with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) CR2 DEFIBRILLATOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key11266797
MDR Text Key230988578
Report Number0003015876-2021-00240
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P170018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date 02/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberCR2
Device Catalogue Number99512-000474
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 01/14/2021
Initial Date FDA Received02/02/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/13/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number3015876-01/14/2021-001-C