Model Number BL5631TD |
Device Problem
Break (1069)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device history records were reviewed and found manufacturing and sterilization records to be acceptable.The investigation is ongoing.
|
|
Event Description
|
The user facility in (b)(6) reported that at the end of the vitesse case, the surgeon was removing the plastic trocar with the trocar forceps when the upper part of the trocar broke resulting in the tube getting stuck in the sclera.The surgeon was able to remove
the tube.There was no patient impact and no additional anesthesia needed.The surgery was increased by four additional minutes.
|
|
Manufacturer Narrative
|
Upon further investigation it was determined that the surgeon removed the tube and it was not stuck in the eye.This event is assessed as not serious and is no longer deemed reportable.Evaluation completed.The device was returned and a visual inspection found the assembly dirty with particulates, and in two pieces.The polyimide sleeve had separated from the hub.A microscopic examination was performed.Remnants of the adhesive could be seen on both the hub and the polyimide sleeve.The sample was sent to the vendor for evaluation.The vendor evaluated the hub and polyimide tube using a light microscope at 20x.It appeared that both the hub and polyimide tube had sufficient adhesive.Additionally, based on the location of the glue on the polyimide tube and hub, it was determined that the hub was properly seated onto the polyimide tube.The device history record review concluded that based on the date of manufacture the referenced lot was produced according to established procedures and all operators involved were trained to the applicable procedures.There were no accompanying non-conformances.The vendor concluded that based on the device history records review and sample evaluation, it is inconclusive as to whether this complaint is manufacturing-related or not.This investigation is complete.
|
|
Manufacturer Narrative
|
The device is in the process of being returned to the manufacturer.The lot history, trend analysis, risk analysis, and/or directions for use review were considered acceptable, with the product performing within anticipated rates.No corrective action required.The investigation is ongoing.
|
|
Search Alerts/Recalls
|