RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Calibration Problem (2890)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2021.Date of report: 02/02/2021.
|
|
Event Description
|
The customer reported the button on the touch screen responded poorly, so calibration was performed but the phenomenon was not improved.The unit was not in use, and there was no patient or user harm reported.
|
|
Manufacturer Narrative
|
G4:01feb2021.B4:25feb2021.H11:g5:k102985.H10: the field service engineer (fse) confirmed the failure through operational checking.The touchscreen calibration did not resolve the issue.The fse replaced the touchscreen, and the unit passed testing.The unit was returned to service.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Event Description
|
The customer reported the button on the touchscreen responded poorly, so calibration was performed but the phenomenon was not improved.The unit was not in use, and there was no patient or user harm reported.
|
|
Manufacturer Narrative
|
The touchscreen assembly was returned to the manufacturer for failure analysis.Visual inspection of the touchscreen assembly revealed no evidence of damage or contamination.The touchscreen was tested and found the ul_lr and ur_ll resistance were out of specification.
|
|
Search Alerts/Recalls
|
|
|