Additional information was added to d9, h3, h4 and h6.H4: the lot was manufactured from july 19, 2020 - july 21, 2020.H10: two (2) actual samples were received for evaluation.During visual inspection, the spike of the first sample was observed attached to the tubing; however, the vent filter was missing.Functional testing was performed on the first sample via a repeater pump system level testing by using sterile water and a leak was observed at the area of the spike where the vent filter was received missing.The reported condition was verified.The cause of the condition could not be determined.During visual inspection, the second sample had detached spike which was not received with the sample.Due to the nature of the second sample, functional testing could not be performed.The reported condition was verified.The cause of the condition could not be determined; however, the most probable cause was due to an inadequate or lack of loctite adhesive being applied in the manufacturing process at the spike connection to the tubing.However, this could not be determined due to the condition of the received sample.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|