Model Number 200H16
Device Problems Calcified (1077); Gradient Increase (1270)
Patient Problems Pulmonary Valve Stenosis (2024); Pulmonary Valve Insufficiency/ Regurgitation (4452)
Event Date 01/11/2021
Event Type  Injury  
Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that 3 years and 5 months post implant of this 16mm pulmonary valved conduit in a (b)(6)-year-old pediatric patient, it was explanted and replaced with an 18mm conduit of the same model.The reason for replacement was reported as conduit calcification with pulmonary stenosis, a peak gradient of 75mmhg, a mean gradient of 42mmhg, and central pulmonary regurgitation.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTEGRA
Type of Device
CONDUIT,VALVED,PULMONIC
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
EI  
091708734
MDR Report Key11278985
MDR Text Key230257437
Report Number2025587-2021-00439
Device Sequence Number1
Product Code MWH
UDI-Device Identifier00643169616912
UDI-Public00643169616912
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H020003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 02/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/08/2020
Device Model Number200H16
Device Catalogue Number200H16
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/12/2021
Initial Date FDA Received02/04/2021
Date Device Manufactured02/08/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age10 YR
Patient Weight36
-
-