ZIMMER BIOMET, INC. UNKNOWN HEAD; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Corroded (1131); Material Erosion (1214); Unintended Movement (3026); Insufficient Information (3190)
Patient Problems Bone Fracture(s) (1870); Failure of Implant (1924); Pain (1994); Ambulation Difficulties (2544); No Information (3190); Metal Related Pathology (4530)
Event Date 05/07/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical devices: item #: unknown, unknown stem lot #: unknown; item #: unknown, unknown liner lot #: unknown; item #: unknown, unknown cup lot #: unknown.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The product location is unknown.Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2021 - 00268.Reported event was unable to be confirmed due to limited information received from the customer.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
Patient's legal counsel reported patient underwent a right hip arthroplasty and was revised seven years later for unknown reasons.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.   medical records were provided and reviewed by a health care professional.Review of the available records identified the following: patient underwent a revision due to metallosis and corrosion, while also experiencing pain a month before revision and unable to bear weight due to a trauma on may 4, 2018.X-ray showed that there was a disassociation of the head from the neck, and metal ions were elevated, with a cobalt level of 1.7mcg/l and a chromium level of 4.1mcg/l.No evidence of infection noted, but found thick black fluid within the hip joint.During the revision it was found the poly liner was damaged, there was a crack on the acetabulum, and corrosion and metallosis was found as well.The stem, head, and liner were revised and new zimmer products put in place.Rom was tested and deemed acceptable prior to closure.The root cause remains unchanged.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported by the patient's legal counsel, patient underwent a right hip arthroplasty.The patient was revised 7 years post implantation due to metallosis and corrosion.The patient began to experience groin pain approximately one month prior to revision and with trauma was unable to bear weight.It was discovered on x-ray there was a disassociation of the femoral head from the neck, and his metal ions were elevated.During revision a linear crack along the rim of the acetabulum was noted as well as a damaged poly liner, corrosion and metallosis.The stem, head and liner were revised.During removal of the stem, an osteotomy was performed as the stem was well integrated, and a fracture occurred and was secured prior to closure.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HEAD
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key11279475
MDR Text Key230287854
Report Number0001822565-2021-00267
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/22/2021
Initial Date FDA Received02/04/2021
Supplement Dates Manufacturer Received09/28/2021
01/13/2022
Supplement Dates FDA Received09/28/2021
02/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexMale
-
-