MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (MOTOR DISORDERS) Back to Search Results
Model Number 37601
Device Problems High impedance (1291); Unintended Collision (1429); Low impedance (2285)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/18/2020
Event Type  malfunction  
Manufacturer Narrative
Country of event: (b)(6).If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that there were high impedances on both sides of the system.Left side pairs c/8, c/9, c/10, 8/11, 9/11, and 10/11 were high in range of 4000-8200 ohms; there was also a low impedance on pair 9/10.Right side pairs c/0, c/1, c/2, 0/3, 1/3 and 2/3 were high in range of 2700-6700 ohms.The patient was programmed with left side contacts 9 & 10 and right side contacts 1 & 2.It was noted the patient reported falling often and that impedance increased during this time.The patient had optimal therapy effect and no shocking sensation.  the patient was implanted for dystonia.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received indicating an x-ray was performed but there was no problem regarding the system integrity observed.No actions were taken, the patient was stable with no side effects that could be due to high impedances.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (MOTOR DISORDERS)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key11280259
MDR Text Key231009454
Report Number3004209178-2021-01987
Device Sequence Number1
Product Code MRU
Combination Product (y/n)N
PMA/PMN Number
H020007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2020
Device Model Number37601
Device Catalogue Number37601
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/01/2021
Initial Date FDA Received02/04/2021
Supplement Dates Manufacturer Received02/10/2021
Supplement Dates FDA Received02/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-