• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES SWITZERLAND SAR NEUROMONITOR BASIC KIT; ICP MICROSENSORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES SWITZERLAND SAR NEUROMONITOR BASIC KIT; ICP MICROSENSORS Back to Search Results
Catalog Number 826631
Device Problem No Device Output (1435)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/19/2021
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported the microsensor could not be zeroed before the procedure.The physician changed the icp monitor and it still could not be zeroed.The microsensor was changed with another of the same microsensor which could be zeroed.There was a 15 minute delay and no adverse consequences reported.
 
Manufacturer Narrative
The microsensor was returned for evaluation: failure analysis - the issue of the complaint was not confirmed: no visible damage to the millar sensor or connector.The catheter received with tape and sutures.The icp express reading passed with 511.The device passed electronic, noise, linearity/hysteresis, and signal drift tests.The root cause of the issue reported by customer could not be determined as the device worked correctly.However, the possible root cause of the defect reported by the customer could be due to incorrect set-up of device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROMONITOR BASIC KIT
Type of Device
ICP MICROSENSORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
MDR Report Key11283314
MDR Text Key234252511
Report Number3013886523-2021-00062
Device Sequence Number1
Product Code GWM
Combination Product (y/n)N
PMA/PMN Number
K914479
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number826631
Device Lot Number4267670
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2021
Initial Date Manufacturer Received 01/19/2021
Initial Date FDA Received02/04/2021
Supplement Dates Manufacturer Received07/08/2021
Supplement Dates FDA Received07/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-