Model Number V60
Device Problem Key or Button Unresponsive/not Working (4063)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(6) 2021.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Event Description
The customer reported that the touch screen is faulty.The customer contacted product support and requested for service repair.The customer reported that the unit was in use on patient, but there was no patient harm reported.Patient information and repair information were requested from the customer, but no response received.
 
Manufacturer Narrative
G4: 01feb2021.B4: 05mar2021.H10: after two non-consecutive visits by philips field service engineer.The equipment was in use, and it was not possible to repair the device because of the covid-19 outbreak situation.The touchscreen material is saved, and preventive maintenance will be used to complete the intervention.If new information is received, a supplemental report will be submitted.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Manufacturer Narrative
G4:05mar2021.B4:12apr2021.The field service engineer (fse) replaced the touchscreen to address the reported issue.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key11286754
MDR Text Key240923300
Report Number2031642-2021-00440
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838009851
UDI-Public(01)00884838009851
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup,Followup
Report Date 01/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Device Catalogue Number1053614
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/08/2021
Initial Date FDA Received02/05/2021
Supplement Dates Manufacturer Received01/08/2021
01/08/2021
Supplement Dates FDA Received03/07/2021
04/12/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/04/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN PATIENT CIRCUIT, MASK, AND HUMIDIFIER.
-
-