It was reported that patient was seen in clinic because of an increase in seizures.Upon interrogation of the device, high impedance of 5391 ohms was detected, which is just over the 5300 ohms trigger for high impedance.Follow-up with the nurse practitioner found that there was no trauma to the vns just prior to the onset of the high impedance.The patient was having increased seizures, but the patient had no known change in perception of vns stimulation.Medications were changed in response to the patient's seizures.X-rays were received and reviewed but that quality of the images were too poor to assess continuity of the leads or complete pin insertion.The manufacturer's programming history database for the patient's generator was reviewed and it was found that the generator was close to 5300 ohms, since a month after explant.Internal investigation identified that a change in the timing of the impedance test may result in higher impedances for model 1000 generators compared to those reported by model 103-106 generators.As indicated in the physician's manual, high lead impedance (>/=5300 ohms), in the absence of other device related complications, is not an indication of a lead or generator malfunction no further relevant information has been received to date.No known relevant surgical intervention has occurred to date.
|