RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Failure to Calibrate (2440); Key or Button Unresponsive/not Working (4063)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2021.Date of report: 05feb2021.
|
|
Event Description
|
The customer called into technical support (ts) reporting that the device¿s touchscreen is not responding to touch.During touchscreen calibration, the unit gave a ¿bad touch detected¿ message and would not calibrate.The customer reported there was no patient involvement at the time the issue was discovered.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed reported problem.The customer replaced the touchscreen to resolve the reported issue.The unit passed required performance verification tests per philips standards.
|
|