• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO SURGICAL SYSTEM INDIA; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO SURGICAL SYSTEM INDIA; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number ROB00036
Device Problems Failure to Power Up (1476); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/15/2021
Event Type  malfunction  
Event Description
It was reported that, during navio machine checking, it was found that no led is activated on the ups after connecting the main power cord.Even after 4 hours of connection the ups batteries did not charge up.The battery is faulty.No patient involved.No other complications were reported.
 
Manufacturer Narrative
Additional information: h4.H3, h6: the navio surgical system (india), product rob00036, (b)(6) intended for use in treatment was not returned for evaluation; thus, a visual and functional evaluation could not be performed, and a relationship between the reported event and the device could not be determined.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.Although the reported problem was not confirmed, a contributing factor noted in the report states that the battery was faulty.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO SURGICAL SYSTEM INDIA
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key11292572
MDR Text Key231116191
Report Number3010266064-2021-00088
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556693506
UDI-Public885556693506
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberROB00036
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/15/2021
Initial Date FDA Received02/08/2021
Supplement Dates Manufacturer Received08/20/2021
Supplement Dates FDA Received08/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNKN NAVIO ROBOTICS INSTR
-
-