AMO MANUFACTURING USA, LLC FS DISPOSABLE INTERFACE; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 590106AN
Device Problem Decrease in Suction (1146)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Type  malfunction  
Manufacturer Narrative
Date of event is unknown as it was not provided.(b)(6).The patient interface (pi) suction ring may lose suction during a procedure.Label copy states corneal fixation vacuum loss can occur.There are several factors that may contribute to suction issues such as doctor¿s technique in applying the suction ring to the cornea, doctor¿s technique in squeezing the pi clip to secure the suction ring to the pi cone and patient anatomy affecting the interface between the patient¿s cornea and the suction ring.If there is any further relevant information, a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that during surgery there was a suction loss while laser was firing.Procedure was completed.No injury for the patient.No further information available.This is 3 of 6 reports.
 
Manufacturer Narrative
Section d9: device available for evaluation: yes date returned to manufacturer: mar 9, 2021 section h3: device returned to manufacturer: yes device evaluation: visual inspection was performed, and no obvious damage/defect was observed.Suction test was performed, and all results were found within specifications.The reported event cannot be confirmed.A record review was performed.A product deficiency review was performed and there is no product deficiency identified.A document, lot history, and trending were reviewed.There is not a recognizable adverse trend.The risks and mitigations associated with the complaint issue are identified in existing risk documents and no new risks were identified as part of this investigation.The review of the device history record (dhr) for the patient interface was performed.The device and its components met all specifications prior to being released.Manufacturing has been ruled out as a potential cause for the reported issue.Based on the investigation results, no corrective action has been issued.Conclusion: based on the information obtained, product malfunction and product deficiency cannot be confirmed.No further investigation is required all pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FS DISPOSABLE INTERFACE
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMO MANUFACTURING USA, LLC
510 cottonwood drive
milpitas CA 95035
MDR Report Key11293796
MDR Text Key230934048
Report Number3006695864-2021-07125
Device Sequence Number1
Product Code HNO
UDI-Device Identifier15050474534688
UDI-Public(01)15050474534688(17)220415(10)60237890
Combination Product (y/n)N
PMA/PMN Number
K060372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 05/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/15/2022
Device Model Number590106AN
Device Catalogue Number590106AN
Device Lot Number60237890
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/10/2021
Initial Date FDA Received02/08/2021
Supplement Dates Manufacturer Received04/19/2021
Supplement Dates FDA Received05/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-