Model Number 71364450
Device Problem Failure to Align (2522)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that, during thr surgery the it was noticed that the threads on ref cup positioner/impactor were not aligning correctly.The procedure was completed without delay using the same device.No other complications were reported.
 
Manufacturer Narrative
Change the product code, catalog number and udi number as per new information received by the initial reporter.
 
Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The threads on the impactor are not aligned correctly, rendering it inoperable.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REF CUP POSITIONER/IMPACTOR
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
brooks rd. 1450
memphis TN
MDR Report Key11298533
MDR Text Key230923020
Report Number1020279-2021-01039
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010607034
UDI-Public03596010607034
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup,Followup
Report Date 06/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71364450
Device Catalogue Number71362299
Device Lot Number18AM14299
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2021
Initial Date Manufacturer Received 01/12/2021
Initial Date FDA Received02/08/2021
Supplement Dates Manufacturer Received02/09/2021
06/01/2021
Supplement Dates FDA Received02/11/2021
06/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-