Ge et al 2020 (evo-d) ¿ ¿eus-guided gastroenterostomy versus enteral stent placement for palliation of malignant gastric outlet obstruction¿.The procedure was performed using a therapeutic gastroscope (gif-xtq160; olympus america, (b)(4)).Once the area of stenosis was reached endoscopically, a standard cannula (tandem xl; boston scientific, (b)(4)) with 0.035 inch guidewire was advanced across the obstruction.Contrast was injected to delineate the length of the stenosis.An appropriately sized uncovered metal duodenal stent (wallflex duodenal; boston scientific, or evolution; cook medical, (b)(4)) was then deployed under direct endoscopic and fluoroscopic guidance across the stenosis." stent ingrowth (16.5% of 97).
|
Pma/510(k) #: k163468.Device evaluation: the duodenal devices involved in this complaint was not returned to cirl for evaluation.With the information provided, a document based investigation will be complete.This file was created from the journal article." ge et al 2019 ".Complaint files (b)(4) (3001845648-2021-00067), (b)(4) (3001845648-2021-00068), (b)(4) (3001845648-2021-00070), (b)(4) (3001845648-2021-00071), (b)(4) (3001845648-2021-00069), (b)(4) (3001845648-2021-00066) and (b)(4) (3001845648-2021-00072) were opened as a result of this paper.This file (b)(4) (3001845648-2021-00067) was opened to investigate stent ingrowth.Lab evaluation: n/a.Documents review including ifu review: as the rpn and lot number of the complaint stents are unknown, a review of the relevant manufacturing records cannot be conducted.However, prior to distribution evo duodenal devices are subjected to a visual inspection and functional checks to ensure device integrity.As the evo duodenal devices from unknown lot number, a review of the relevant manufacturing records cannot be conducted.As per the instructions for use, ifu0053-10, which accompanies this device it informs the user about the potential complications "additional complications include, but are not limited to: tumour ingrowth or overgrowth".There is no evidence to suggest that the customer did not follow the instructions for use.Root cause review: a definitive root cause could not be determined from the available information.A possible root cause could be attributed to patient condition related, as per instructions for use, tumour ingrowth or overgrowth is listed as a complication following the use of this device.Summary: complaint is confirmed based on the customers testimony.As per medical advisor opinion require intervention/additional procedures s=4.Procedure related complication.It is covered in ifu by tumour ingrowth or overgrowth.Complaints of this nature will continue to be monitored for potential emerging trends.
|