Analysis of the electronic data files from aed serial number (b)(4) identified that the aed entered into service on may 06, 2020, but was not set up per the device user instructions; specifically, when the aed was placed into service, the pads were not connected to the aed.Once in service, the aed detected that the pads were not connected and attempted to alert the user by flashing its red asi and chirping.The aed continued to flash its red asi and chirp, without any evidence of human interaction to address the unit's condition, until (b)(6) 2020, when the aed's battery pack became fully depleted.Approximately four weeks later, on (b)(6) 2021, the aed was deployed for the rescue attempt but would not power on due to the depleted battery pack.
|