BOSTON SCIENTIFIC NEUROMODULATION VERCISE CARTESIA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
|
Back to Search Results |
|
Model Number DB-2202-45 |
Device Problems
Fracture (1260); High impedance (1291)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/19/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional suspect medical device components involved in the event: product family: dbs-linear leads, upn: (b)(4), model: db-2202-45, serial: (b)(4), batch: 7075773.
|
|
Event Description
|
It was reported that the during a programming appointment high impedances were observed on the patient's leads.The physician, suspecting lead fracture, ordered x-rays to confirm lead fracture.
|
|
Event Description
|
It was reported that the during a programming appointment high impedances were observed on the patient's leads.The physician, suspecting lead fracture, ordered x-rays to confirm lead fracture.Additional information was received that the x-ray did not show a visible fracture.The patient was seen in the clinic and it was determined that no further action will be taken at this time.The lead contacts suggesting the fracture are not being used for the stimulation.
|
|