BOSTON SCIENTIFIC NEUROMODULATION VERCISE CARTESIA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS Back to Search Results
Model Number DB-2202-45
Device Problems Fracture (1260); High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2021
Event Type  malfunction  
Manufacturer Narrative
Additional suspect medical device components involved in the event: product family: dbs-linear leads, upn: (b)(4), model: db-2202-45, serial: (b)(4), batch: 7075773.
 
Event Description
It was reported that the during a programming appointment high impedances were observed on the patient's leads.The physician, suspecting lead fracture, ordered x-rays to confirm lead fracture.
 
Event Description
It was reported that the during a programming appointment high impedances were observed on the patient's leads.The physician, suspecting lead fracture, ordered x-rays to confirm lead fracture.Additional information was received that the x-ray did not show a visible fracture.The patient was seen in the clinic and it was determined that no further action will be taken at this time.The lead contacts suggesting the fracture are not being used for the stimulation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERCISE CARTESIA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
MDR Report Key11307967
MDR Text Key231305308
Report Number3006630150-2021-00378
Device Sequence Number1
Product Code NHL
UDI-Device Identifier08714729905288
UDI-Public08714729905288
Combination Product (y/n)N
PMA/PMN Number
P150031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/18/2022
Device Model NumberDB-2202-45
Device Catalogue NumberDB-2202-45
Device Lot Number7075727
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/19/2021
Initial Date FDA Received02/10/2021
Supplement Dates Manufacturer Received02/25/2021
Supplement Dates FDA Received03/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age54 YR
-
-