• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PERSONA TIBIAL ARTICULAR SURFACE PROVISIONAL; PROSTHESIS KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PERSONA TIBIAL ARTICULAR SURFACE PROVISIONAL; PROSTHESIS KNEE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/12/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint device has been returned, but the device investigation has not yet been completed.Once the evaluation is completed, a supplemental medwatch will be submitted.
 
Event Description
It was reported that the device was fractured after surgery during sterilization process.Attempt for further information has been made, but no further information has been provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated.Visual evaluation of the returned device shows signs of repeated use and the post has fractured off.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.A previous investigation into this issue determined that the likely root cause for the tasp fractures is bending/torsional loading on the device.A notification was sent to the field on august 7th, 2014 to provide additional clarifications relating to the instructions for use of the tasp construct for all persona users on file at the time of the field notice.Ps tasp top components and standard (non-cps) tasp bottom components have been modified to prevent fracture.The fractured tasp component in this complaint was manufactured before the design modification.The root cause is considered to be a previously addressed design issue.The event is being reviewed through the capa process.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA TIBIAL ARTICULAR SURFACE PROVISIONAL
Type of Device
PROSTHESIS KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11308804
MDR Text Key232453368
Report Number0001822565-2021-00355
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
PMA/PMN Number
K172524
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42527000707
Device Lot Number62796252
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/19/2021
Initial Date FDA Received02/10/2021
Supplement Dates Manufacturer Received03/17/2021
Supplement Dates FDA Received03/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-2578-2014
Patient Sequence Number1
-
-