The device was returned to olympus for evaluation.The evaluation confirmed the user report.Leaking was present due to a cut at the a-rubber (bending section).The bending section support pin was lifted and the scope had fluid invasion from the leak.As part of the investigation, a review of the device history record (dhr) and a review of the instructions for use (ifu) were also conducted.The review of the dhr did not find any abnormalities or anomalies identified during production.The device met all specifications upon release.The ifu contains the following statements: "precautions: perform a leakage test on the endoscope after each precleaning procedure.Do not use the endoscope if a leak is detected.Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions.Use of a leaking endoscope may also pose an infection control risk." "do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist." "do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged." the exact root cause could not be established.A likely cause is when the user inserts the subject device into the lower kidney calyx or urinary tract, the endoscope is forcefully pushed through with the bending section being bent.
|