Catalog Number 256082 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problems
No Information (3190); Insufficient Information (4580)
|
Event Date 01/20/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Eua # (b)(4).Medical device expiration date: unknown.Initial reporter state: address information was not able to be obtained, however, (b)(6) was used as a place holder.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown (b)(4).
|
|
Event Description
|
It was reported while testing for sars cov-2 potential false positive results were obtained.The information source was from a facebook post comment.No follow-up was performed as contact information was not provided.Eua # (b)(4).
|
|
Manufacturer Narrative
|
H.6.Investigation: bd has received several customer complaints for false positive results, when using bd sars-cov-2 reagents for bd veritor¿ system.The current investigation concerns multiple lots of bd sars-cov-2 reagents for bd veritor¿ system.Bd takes a systematic approach to investigating false positive complaints that are received.This approach involves review of manufacturing batch history records, testing of retention samples, and testing of customer returned samples if applicable.The investigation did not find a root cause for the false positive results that were observed.Bd cannot confirm the complaint based on the investigation that was performed.A corrective and preventive action (capa#1878253) is already initiated to investigate the root cause and some mitigation actions are already being addressed.See h.10.
|
|
Event Description
|
It was reported while testing for sars cov-2 potential false positive results were obtained.The information source was from a facebook post comment.No follow-up was performed as contact information was not provided.Eua # (b)(4).
|
|