MAUDE Adverse Event Report: OGDEN MANUFACTURING PLANT OPTIFLUX 180NRE DIALYZER FINISHED ASSY.; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OGDEN MANUFACTURING PLANT OPTIFLUX 180NRE DIALYZER FINISHED ASSY.; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number 0500318E
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/29/2021
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: the plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event Description
A user facility registered nurse (rn) reported that a dialyzer blood leak occurred immediately after the initiation of the patient¿s hemodialysis (hd) treatment.The blood leak was noted as being an internal blood leak that could be visually observed within the dialysate.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.Blood leak test strips were used and tested positive for the presence of blood.There was no defect or damage seen on the dialyzer.The patient¿s estimated blood loss (ebl) was approximately 250 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The patient was restarted on the machine and treatment completed successfully with new supplies.The complaint device was reported to be unavailable to be returned to the manufacturer for evaluation.
 
Manufacturer Narrative
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLUX 180NRE DIALYZER FINISHED ASSY.
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
MDR Report Key11309090
MDR Text Key247026197
Report Number1713747-2021-00043
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861100156
UDI-Public00840861100156
Combination Product (y/n)N
PMA/PMN Number
K002761
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Model Number0500318E
Device Catalogue Number0500318E
Device Lot Number20PU02017
Was Device Available for Evaluation? No
Device AgeMO
Initial Date Manufacturer Received 01/29/2021
Initial Date FDA Received02/10/2021
Supplement Dates Manufacturer Received02/14/2021
Supplement Dates FDA Received02/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
FRESENIUS 2008T MACHINE; FRESENIUS BLOODLINES
Patient Age81 YR
Patient Weight76
-
-