• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO MANUFACTURING USA, LLC UNK_PI-RET_PATIENT INTERFACE; KERATOME, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING USA, LLC UNK_PI-RET_PATIENT INTERFACE; KERATOME, AC-POWERED Back to Search Results
Model Number 590106AN
Device Problem Decrease in Suction (1146)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/08/2021
Event Type  malfunction  
Manufacturer Narrative
The patient interface (pi) suction ring may lose suction during a procedure.Label copy states corneal fixation vacuum loss can occur.There are several factors that may contribute to suction issues such as doctor¿s technique in applying the suction ring to the cornea, doctor¿s technique in squeezing the pi clip to secure the suction ring to the pi cone and patient anatomy affecting the interface between the patient¿s cornea and the suction ring.A review of records related to the device including labeling, complaint trending, and risk documentation will be performed.Upon completion of this review, if there is any further relevant information obtained, a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that a suction loss after the laser fired occurred with a patient interface with an unknown lot number.The surgery was completed by switching out the pi.No patient injury was reported.This is for the first patient interface #2 of 2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK_PI-RET_PATIENT INTERFACE
Type of Device
KERATOME, AC-POWERED
Manufacturer (Section D)
AMO MANUFACTURING USA, LLC
510 cottonwood drive
milpitas CA 95035
Manufacturer Contact
somyata nagpal
510 cottonwood drive
milpitas, CA 95035
7142478552
MDR Report Key11313699
MDR Text Key231593163
Report Number3006695864-2021-07087
Device Sequence Number1
Product Code HNO
UDI-Device Identifier15050474534688
UDI-Public(01)15050474534688
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K060372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Physician
Type of Report Initial
Report Date 02/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number590106AN
Device Catalogue Number590106AN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/20/2021
Initial Date FDA Received02/11/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-