• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD BBL CDC ANAEROBE 5% SHEEP BLOOD AGAR; CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD BBL CDC ANAEROBE 5% SHEEP BLOOD AGAR; CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL Back to Search Results
Model Number 221734
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/20/2021
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
 
Event Description
It was reported that prior to use with bd bbl¿ cdc anaerobe 5% sheep blood agar mold contamination was discovered.
 
Manufacturer Narrative
Investigation summary: during manufacturing of material 221734, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck (2 to 8 degrees c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.The batch history record for batch 0289033 was satisfactory at time of release and no quality notifications were generated during manufacturing and inspection.The release testing that is performed on this product does include bioburden testing.A sample of plates are incubated at 25 degrees c and at 35 degrees c for approximately 72 hours.All bioburden testing performed on this batch was satisfactory per bd internal procedures.Affected product does not have any sterility claims; the product is tested for bioburden prior to release to ensure that it conforms to product specifications.However, this does not ensure that the end-user will not receive a contaminated plate.The complaint history was reviewed, and no other complaints have been taken on batch 0289033.Retention samples from batch 0289033 were not available for inspection.One photo was received for investigation.The photo shows three sleeves from batch 0289033 with microbial growth visible in the bottom four plates of one sleeve.No returns were received for investigation.This complaint can be confirmed for contamination.Bd will continue to trend complaints for broken plates and contamination.Based on the low defect rate for this batch, no actions are planned at this time.
 
Event Description
It was reported that prior to use with bd bbl¿ cdc anaerobe 5% sheep blood agar mold contamination was discovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BBL CDC ANAEROBE 5% SHEEP BLOOD AGAR
Type of Device
CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
MDR Report Key11318129
MDR Text Key242460613
Report Number1119779-2021-00295
Device Sequence Number1
Product Code JSG
UDI-Device Identifier10382902217340
UDI-Public10382902217340
Combination Product (y/n)N
PMA/PMN Number
K803025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 09/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date01/28/2021
Device Model Number221734
Device Catalogue Number221734
Device Lot Number0289033
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/22/2021
Initial Date FDA Received02/11/2021
Supplement Dates Manufacturer Received09/01/2021
Supplement Dates FDA Received09/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-