BECTON, DICKINSON & CO. (SPARKS) BD BBL CDC ANAEROBE 5% SHEEP BLOOD AGAR; CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL
|
Back to Search Results |
|
Model Number 221734 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problems
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/20/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
|
|
Event Description
|
It was reported that prior to use with bd bbl¿ cdc anaerobe 5% sheep blood agar mold contamination was discovered.
|
|
Manufacturer Narrative
|
Investigation summary: during manufacturing of material 221734, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck (2 to 8 degrees c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.The batch history record for batch 0289033 was satisfactory at time of release and no quality notifications were generated during manufacturing and inspection.The release testing that is performed on this product does include bioburden testing.A sample of plates are incubated at 25 degrees c and at 35 degrees c for approximately 72 hours.All bioburden testing performed on this batch was satisfactory per bd internal procedures.Affected product does not have any sterility claims; the product is tested for bioburden prior to release to ensure that it conforms to product specifications.However, this does not ensure that the end-user will not receive a contaminated plate.The complaint history was reviewed, and no other complaints have been taken on batch 0289033.Retention samples from batch 0289033 were not available for inspection.One photo was received for investigation.The photo shows three sleeves from batch 0289033 with microbial growth visible in the bottom four plates of one sleeve.No returns were received for investigation.This complaint can be confirmed for contamination.Bd will continue to trend complaints for broken plates and contamination.Based on the low defect rate for this batch, no actions are planned at this time.
|
|
Event Description
|
It was reported that prior to use with bd bbl¿ cdc anaerobe 5% sheep blood agar mold contamination was discovered.
|
|