|
Model Number REPLY VDR |
Device Problems
Failure to Capture (1081); Over-Sensing (1438); Failure to Sense (1559); Low impedance (2285)
|
Patient Problems
No Information (3190); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/14/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device involved in this mdr report is not approved in the united states; however, it is similar to a device manufactured by microport crm that was cleared or approved by fda for marketing in the united states.
|
|
Event Description
|
The pacing system was implanted on (b)(6) 2019 to a patient with a complete av block.Reportedly, during a follow-up performed in (b)(6) 2020, a decrease of the lead impedance to 205 ohms was observed.Episodes showing noise on the atrial channel were recorded in the device memory.The pacing mode regularly switched in vvi.In addition, ventricular events were not detected.Though the sensitivity was programmed to the maximum value, failure to sense was observed.The capture threshold was measured at 1.0 v.X-ray images did not reveal any lead issue or dislodgement.A re-intervention was scheduled to check the pacing system.When unscrewing the leads during a re-intervention performed in (b)(6) 2020, blood traces were observed in both connector cavities.After cleaning the traces of blood, pacing system analyzer (psa) measurements were performed for the atrioventricular lead but no anomaly was found.The ventricular measurements showed a good r-wave detection (5-8 mv), lead impedance measurements were within normal range (350-400 ohms) and the ventricular capture threshold was at 0,6v/0,4ms.The atrial measurements showed a good p-wave detection (0,5-0,6 mv).Therefore, the lead was re-connected to the pacemaker and re-inserted in the pocket.During a follow-up performed on (b)(6) 2021, the same issue was observed.The lead impedance decreased to 215 ohms and loss of atrial capture was observed.The pacing mode regularly switched in vvi.However, ventricular events were not detected, even though the patient escape rhythm was 30 bpm.A lead and/or connection issue were suspected by the physician, who scheduled a re-intervention.On (b)(6) 2021, the pacemaker was replaced, and the lead was abandoned in patient¿s body.Preliminary analysis revealed that the observed issue most probably resulted from an atrioventricular lead issue and/or lead-pacemaker connection issues at atrial and ventricular levels.
|
|
Manufacturer Narrative
|
Please refer to the attached analysis report.
|
|
Event Description
|
The pacing system was implanted on (b)(6) 2019 to a patient with a complete av block.Reportedly, during a follow-up performed in (b)(6) 2020, a decrease of the lead impedance to 205 ohms was observed.Episodes showing noise on the atrial channel were recorded in the device memory.The pacing mode regularly switched in vvi.In addition, ventricular events were not detected.Though the sensitivity was programmed to the maximum value, failure to sense was observed.The capture threshold was measured at 1.0 v.X-ray images did not reveal any lead issue or dislodgement.A re-intervention was scheduled to check the pacing system.When unscrewing the leads during a re-intervention performed in (b)(6) 2020, blood traces were observed in both connector cavities.After cleaning the traces of blood, pacing system analyzer (psa) measurements were performed for the atrioventricular lead but no anomaly was found.The ventricular measurements showed a good r-wave detection (5-8 mv), lead impedance measurements were within normal range (350-400 ohms) and the ventricular capture threshold was at 0,6v/0,4ms.The atrial measurements showed a good p-wave detection (0,5-0,6 mv).Therefore, the lead was re-connected to the pacemaker and re-inserted in the pocket.During a follow-up performed on (b)(6) 2021, the same issue was observed.The lead impedance decreased to 215 ohms and loss of atrial capture was observed.The pacing mode regularly switched in vvi.However, ventricular events were not detected, even though the patient escape rhythm was 30 bpm.A lead and/or connection issue were suspected by the physician, who scheduled a re-intervention.On (b)(6) 2021, the pacemaker was replaced, and the lead was abandoned in patient's body.Preliminary analysis revealed that the observed issue most probably resulted from an atrioventricular lead issue and/or lead-pacemaker connection issues at atrial and ventricular levels.
|
|
Search Alerts/Recalls
|
|
|