CONCORD MANUFACTURING 2008T HEMODIALYSIS SYS., WITH CDX; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
|
Back to Search Results |
|
Model Number 190713 |
Device Problems
Thermal Decomposition of Device (1071); Melted (1385); Smoking (1585)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/28/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Plant investigation: the plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.¿.
|
|
Event Description
|
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t hemodialysis (hd) machine had a burned power supply and a melted fan cable.A patient care technician (pct) was disinfecting the machine when bleach spilled causing the fan cable to melt and fuses on the power supply to blow.The biomed also stated that a puff of smoke was observed upon powering down the machine.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.The machine has approximately 32,000 hours and the parts are the original fresenius parts on the machine.The biomed reported that there was no damage observed on any other components, or any other additional issues, associated with the burned power supply and the melted fan cable the biomed replaced the power supply and the fan cable, which resolved the issue.The unit was returned to service at the user facility without issue and without reoccurrence of the event.The power supply and the fan cable were discarded by the biomed and are not available to be returned to the manufacturer for physical evaluation.
|
|
Event Description
|
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t hemodialysis (hd) machine had a burned power supply and a melted fan cable.A patient care technician (pct) was disinfecting the machine when bleach spilled causing the fan cable to melt and fuses on the power supply to blow.The biomed also stated that a puff of smoke was observed upon powering down the machine.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.The machine has approximately 32,000 hours and the parts are the original fresenius parts on the machine.The biomed reported that there was no damage observed on any other components, or any other additional issues, associated with the burned power supply and the melted fan cable the biomed replaced the power supply and the fan cable, which resolved the issue.The unit was returned to service at the user facility without issue and without reoccurrence of the event.The power supply and the fan cable were discarded by the biomed and are not available to be returned to the manufacturer for physical evaluation.
|
|
Manufacturer Narrative
|
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius field service technician (fst).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
|
|