RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Use of Device Problem (1670); Defective Component (2292)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2021.Date of report: 15feb2021.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Event Description
|
The customer called into technical support (ts) reporting that the device¿s nav-ring is not working and unable to adjust the settings.Requesting onsite service for nav-ring replacement.The customer reported there was no patient involvement at the time the issue was discovered.
|
|
Manufacturer Narrative
|
G4:22mar2021.B4:02apr2021.The manufacturer's field service engineer (fse) was dispatched to the site and confirmed the reported problem of the nav-ring not working and unable to adjust the settings.The fse replaced the front bezel to return the device back to working condition.The device passed all tests successfully and was returned unit to service.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Manufacturer Narrative
|
H11: it is unknown if the unit was in inpatient use at the time of the reported issue; however, no patient or user harm was reported.H10: the front bezel was returned back for failure investigation, and it was identified that previous investigations have shown liquid ingress is due to the variability of the assembly process causing the reported navigation ring failure.
|
|