• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC. FLUENT FLUID MANAGEMENT SYSTEM; INSUFFLATOR, HYSTEROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC. FLUENT FLUID MANAGEMENT SYSTEM; INSUFFLATOR, HYSTEROSCOPIC Back to Search Results
Model Number FLT-112S
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/02/2021
Event Type  malfunction  
Manufacturer Narrative
Device history record (dhr) review was conducted for the reported identification number.The lot was released meeting all qa specifications.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
 
Event Description
It was reported that during a fluent procedure, the tissue sock and cannister started bubbling and overfilled.The surgeon decided to abort the case and change to a dilatation and curetage procedure.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENT FLUID MANAGEMENT SYSTEM
Type of Device
INSUFFLATOR, HYSTEROSCOPIC
Manufacturer (Section D)
HOLOGIC, INC.
250 campus drive
marlborough, MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough, MA 01752
Manufacturer Contact
daniel guevara
562 parkway
coyol free zone building b24
san jos[?], alajuela 20102- CRI
CS   20102 CRI
MDR Report Key11330090
MDR Text Key232055436
Report Number1222780-2021-00012
Device Sequence Number1
Product Code HIG
UDI-Device Identifier15420045507401
UDI-Public(01)15420045507401(17)230929(10)20J29RJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K180825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/29/2023
Device Model NumberFLT-112S
Device Catalogue NumberFLT-112S
Device Lot Number20J29RJ
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/02/2021
Initial Date FDA Received02/15/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/29/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-