The tempus ls-manual was used on a apnoeic, pulseless non-moving patient.The customer advised that the pads were plugged into the device which displayed 'check defibrillator electrodes'.At some point during movement of the patient, the crew noticed the electrode cable for the device was disconnected and on inspection noticed the plug in portion of the cable was broken and a piece remains stuck inside the device.The customer has advised that the tempus ls-manual was dropped during the patient transport.The unit was checked back at base by senior facility staff and the identified issue could not be replicated.
|