OLYMPUS MEDICAL SYSTEMS CORP. OER-3 100V; ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-3
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/21/2021
Event Type  malfunction  
Manufacturer Narrative
The air/water/instrument channel connector of the oer-3 was returned to olympus medical systems corp.(omsc) for evaluation.Omsc confirmed that the air/water/instrument channel connector was broken.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.The exact cause was unknown; however, the following was supposed to be the cause.The user repeatedly applied a rotational force to the connector of the reprocessing basin of the device.
 
Event Description
Olympus was informed from the facility that in unspecified timing the air/water/instrument channel connector of the reprocessing basin of the device was broken and a local service engineer repaired the device at the facility.Other detailed information was not provided.There was no report of patient injury associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OER-3 100V
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11330487
MDR Text Key233373678
Report Number8010047-2021-02650
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial
Report Date 02/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/21/2021
Initial Date FDA Received02/15/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-