Model Number 788426
Device Problem Biocompatibility (2886)
Patient Problems No Code Available (3191); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 12/05/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the patient was implanted with left ureteral stent due to severe left hydronephrosis.Later, a large number of calculi was attached to the stent tube soon after the stent was placed.The warranty period was one year.
 
Manufacturer Narrative
The reported event was inconclusive.It was unknown whether the device had met specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.No sample was returned for evaluation.A potential root cause for this failure could be due to "material selection".The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "indications for use: the inlay optima® ureteral stent and multi-length ureteral stent with suture are indicated to relieve obstruction in a variety of benign, malignant and post-traumatic conditions in the ureter.These conditions include stones and/or stone fragments, or other ureteral obstructions such as those associated with ureteral stricture, malignancy of abdominal organs, retroperitoneal fibrosis or ureteral trauma, or in association with extracorporeal shock wave lithotripsy (eswl).The stent may be placed using endoscopic surgical techniques or percutaneously using standard radiographic technique.It is recommended that the indwelling time not exceed 365 days.The stent is not intended as a permanent indwelling device.Description: the inlay optima® ureteral stent and multi-length ureteral stent is a coated, double pigtail ureteral stent with a monofilament suture loop attached to aid in stent removal.The stent is available in two forms: a single size or a customizable multi-length size.The following items are included with each stent: 1 ureteral stent with suture 1 push catheter with radiopaque band 1 pigtail straightener 1 guidewire* (optional) *note: a 4.7 fr stent is compatible with a.035¿ guidewire and 6, 7, and 8 fr stents are compatible with a.038¿ guidewire.In vitro testing conducted on the inlay optima® ureteral stent and multi-length ureteral stent indicate reduced accumulation of urine calcium salts as assayed by calcium when compared to a control.Correlation of in vitro data to clinical outcome has not been established.Choong, sks, wood, s, whitfield, hn.¿a model to quantify encrustation on ureteric stents, urethral catheters, and polymers intended for urological use,¿ bju international (2000), 86, 414-421.Contraindications: no known contraindications for use.Precautions: ¿ suture may be cut off prior to stent placement.Remove suture if indwelling time is expected to be longer than 14 days.¿ avoid improper handling of stent such as bending, kinking, tearing, etc.Misuse could damage the overall integrity of the stent.¿ ureteral stents should be checked periodically for signs of encrustation and proper function.Periodic checks of the stent by cystoscopic and/or radiographic procedures are recommended at intervals deemed to be appropriate by the physician in consideration of the individual patient¿s condition and other patient specific factors.When long-term use is indicated, it is recommended that indwelling time not exceed 365 days.The stent is not intended as a permanent indwelling device.* ¿ with any ureteral stent, migration is a possible complication, which could require medical intervention for removal.Selection of too short a stent may result in migration.¿ care should be exercised when removing the stent from the inner polybag to eliminate tearing or fragmentation.¿ the insertion of a ureteral stent should only be done by those individuals." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
 
Event Description
It was reported that the stent was placed to alleviate the hydronephrosis for a patient who was treated for severe left hydronephrosis for unknown reasons.It was reported that a large number of calculi attached to the stent tube formed soon after the indwelling.The warranty period is one year.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD INLAY OPTIMA URETERAL STENT
Type of Device
BARD URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key11330935
MDR Text Key243574772
Report Number1018233-2021-00521
Device Sequence Number1
Product Code FAD
UDI-Device Identifier00801741015687
UDI-Public(01)00801741015687
Combination Product (y/n)N
PMA/PMN Number
K043193
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 05/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number788426
Device Catalogue Number788426
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/19/2021
Initial Date FDA Received02/16/2021
Supplement Dates Manufacturer Received05/13/2021
Supplement Dates FDA Received06/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-