• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION OXIRIS; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION OXIRIS; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955503
Device Problems Disconnection (1171); Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Oxiris s has been temporarily approved for use in the us under emergency use authorization (b)(4) with a specific indication to treat patients with covid-19 infection.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that approximately one hour into continuous renal replacement therapy using a prismax and a prismaflex oxiris set, a split and separated tubing occurred.This resulted to an external blood leak.The amount of blood loss was not reported.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.The visual inspection of the provided picture showed an external blood leak at the level of the replacement pump segment and that the replacement pump segment was observed to be disconnected from the support plate on one site.The reported condition was verified.The cause was manufacturing related.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXIRIS
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11331332
MDR Text Key233453955
Report Number8010182-2021-00051
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955503
Device Lot Number20D2710Z
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/25/2021
Initial Date FDA Received02/16/2021
Supplement Dates Manufacturer Received03/15/2021
Supplement Dates FDA Received03/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PRISMAX MACHINE
-
-