Catalog Number 114745L
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 01/26/2021
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone number:  (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that upon opening the packaging of a revaclear 300, two dark spots were observed inside the fibers.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: the actual sample was not received for evaluation, however, a picture was provided.Visual inspection observed an image of the header cap area of a dialyzer.On the top of the fiber bundle, were two small dark 'spots', that appeared to be possible singe or burn marks.While the spots looked to be fixed to the bundle top, the actual sample would have to be analyzed for confirmation.The discoloration may be consistent with singe or burn marks caused by debris on the hot blades, the manufacturing equipment used to cut the fiber bundle.Fixed particulate matter, that is not loose in the header cap area, may have potential to enter the dialysate path, however would not be probable to enter the blood path.The reported condition was verified.The cause could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVACLEAR 300
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11331423
MDR Text Key241931255
Report Number3006552611-2021-00005
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414123055
UDI-Public(01)07332414123055
Combination Product (y/n)Y
PMA/PMN Number
K130039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/20/2023
Device Catalogue Number114745L
Device Lot NumberC620310902
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/26/2021
Initial Date FDA Received02/16/2021
Supplement Dates Manufacturer Received03/16/2021
Supplement Dates FDA Received03/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-