ROCHE DIAGNOSTICS COAGUCHEK XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Catalog Number 03666794001
Device Problem Output Problem (3005)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/21/2021
Event Type  malfunction  
Manufacturer Narrative
Unique identifier (udi) (b)(4).Occupation is lay user/patient the reporter's meter was provided for investigation where it was tested using retention strips and retention controls.Testing results (qc range = 4.1 - 6.8 inr): qc 1: 5.1 inr qc 2: 5.2 inr qc 3: 5.3 inr the obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.The meter's error memory was reviewed and no error 5 or error 7 messages were observed at the same time as a result.No other errors were observed.The control testing was saved in the meter memory correctly.A full display check was performed and observed no missing segments or display issues.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Event Description
The initial reporter complained of a display issue with coaguchek xs meter serial number(b)(4).Customer stated they initially received an error 5 message on the meter during a test.Error 5 can indicate not enough blood was applied to the test strip.Customer tested again and received a result of 5.2 inr and the customer stated this result then switched to an error 7 on the display.No segments were missing from the meter display.The allegation could lead to misinterpretation of results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key11331935
MDR Text Key255006730
Report Number1823260-2021-00475
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 02/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number03666794001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2021
Initial Date Manufacturer Received 01/21/2021
Initial Date FDA Received02/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-