Catalog Number 03666794001 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/21/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Unique identifier (udi) (b)(4).Occupation is lay user/patient the reporter's meter was provided for investigation where it was tested using retention strips and retention controls.Testing results (qc range = 4.1 - 6.8 inr): qc 1: 5.1 inr qc 2: 5.2 inr qc 3: 5.3 inr the obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.The meter's error memory was reviewed and no error 5 or error 7 messages were observed at the same time as a result.No other errors were observed.The control testing was saved in the meter memory correctly.A full display check was performed and observed no missing segments or display issues.The investigation did not identify a product problem.The cause of the event could not be determined.
|
|
Event Description
|
The initial reporter complained of a display issue with coaguchek xs meter serial number(b)(4).Customer stated they initially received an error 5 message on the meter during a test.Error 5 can indicate not enough blood was applied to the test strip.Customer tested again and received a result of 5.2 inr and the customer stated this result then switched to an error 7 on the display.No segments were missing from the meter display.The allegation could lead to misinterpretation of results.
|
|