MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number J2C1711
Device Problem Material Rupture (1546)
Patient Problem No Patient Involvement (2645)
Event Date 01/22/2021
Event Type  malfunction  
Manufacturer Narrative
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the bladder of a small volume infusor ruptured.The bladder ruptured occurred while filling the device with medication prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h4 and h6.H4: the lot was manufactured between august 28, 2020 - august 31, 2020.H10: the device was received for evaluation.Visual inspection on the returned unit via the naked eye noted the bladder has been ruptured.The ruptured bladder was examined for signs of abnormality that could have caused the rupture and no issues were noted.The reported condition was verified.The cause of the condition could not be determined; however, the most probable cause of the condition is manufacturing related.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11334041
MDR Text Key245938170
Report Number1416980-2021-00677
Device Sequence Number1
Product Code MEB
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberJ2C1711
Device Lot Number20H045
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2021
Initial Date Manufacturer Received 01/22/2021
Initial Date FDA Received02/16/2021
Supplement Dates Manufacturer Received02/21/2021
Supplement Dates FDA Received03/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-