LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problem Electrical Power Problem (2925)
Patient Problem No Patient Involvement (2645)
Event Date 01/21/2021
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the heater-cooler system (b)(4).The incident occurred in (b)(6).The technician in charge of the maintenance traced the fault back to a defective power supply.The part was replaced and the problem solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.
 
Event Description
Livanova (b)(4) received a report of a heater-cooler system (b)(4) power failure during maintenance.There was no patient involvement.
 
Manufacturer Narrative
Reportability decision changed to not reportable event since, through follow up communication, livanova learned that the reported error was on cardioplegia side which is not likely to result in death or serious injury.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindbergstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key11337083
MDR Text Key243922796
Report Number9611109-2021-00084
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/21/2021
Initial Date FDA Received02/17/2021
Supplement Dates Manufacturer Received02/23/2021
Supplement Dates FDA Received03/16/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/16/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-