OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/27/2021
Event Type  malfunction  
Manufacturer Narrative
The subject device has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during a repair at olympus (b)(4), it was found that when the subject device was turned on the indicators on the front panel of the subject device were not lit up and the alarm sounds were generated.This phenomenon was occurred after new main circuit board was installed into the subject device, but another main circuit board was installed, the issue did not occur.There was no abnormality on the exterior of the main circuit board.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the subject device evaluation result.The subject device was returned to olympus china.The main circuit board of the subject device was returned to olympus medical systems corp.(omsc) for investigation.Omsc reviewed the device history record (dhr) of the subject device and confirmed no irregularity.Based on the investigation, this phenomenon was attributed to the failure of the pressure sensor mounted on the main circuit board.The cause of the failure of the pressure sensor mounted on the main circuit board is presumed to be due to any of the following process errors.- oxygen entered the device and the electrode of the pressure sensor was oxidized and corroded.The wiring inside the pressure sensor was peeled off due to the oxidation corrosion.- because foreign matter (epoxy) had adhered to the mounting of the component due to a mistake, the wires inside the pressure-sensor had peeled off due to adherence of the foreign matter (epoxy).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11337228
MDR Text Key232798995
Report Number8010047-2021-02746
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/27/2021
Initial Date FDA Received02/17/2021
Supplement Dates Manufacturer Received03/26/2021
Supplement Dates FDA Received04/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-