• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS NORTH AMERICA LLC BIPAP MACHINE; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS NORTH AMERICA LLC BIPAP MACHINE; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problems Device Alarm System (1012); Device Displays Incorrect Message (2591); Power Problem (3010); Therapeutic or Diagnostic Output Failure (3023)
Patient Problem Discomfort (2330)
Event Date 02/09/2021
Event Type  malfunction  
Event Description
A covid-19 positive critically ill pt was on a phillips v60 bipap machine that high priority alarm triggered, with the following alarms showing "check vent: ventilator restarted, check vent backup alarm failed, check vent aux supply failed, check 35v supply failed, low inspiratory pressure, oxygen not available." the pt was removed from the machine, supplemental oxygen used until pt was emergently intubated a few minutes later.The failure is from a power management (pcba) board failure in the machine and the unit involved is part of a philips urgent medical device correction alert (02/2020 and 08/2020).These unit and 12 others (13 total) have been taken out of service until philips can replace the pcba's.Philips was contacted on 02/10/2020 to the expedite replacement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP MACHINE
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
MDR Report Key11337596
MDR Text Key232429356
Report NumberMW5099425
Device Sequence Number1
Product Code MNT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Type of Report Initial
Report Date 02/12/2021
13 Devices were Involved in the Event: 1   2   3   4   5   6   7   8   9   10   11   12   13  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/16/2021
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age45 YR
-
-