• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPIC CO2 REGULATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPIC CO2 REGULATION UNIT Back to Search Results
Model Number UCR
Device Problem Backflow (1064)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Since the subject device has been not returned to omsc for the evaluation, the exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at olympus service operation repair center (sorc), it was found that there was a trace of liquid ingress into the inside tube of the subject device.There was no report of patient injury associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Since the subject device was not returned to olympus medical systems corp.(omsc), it could not be investigated.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.However, based on the information from olympus service operation repair center (sorc), there was the possibility that this phenomenon was attributed to the backflow of liquid because there was a trace of liquid ingress into the inside tube of the subject device.If additional information becomes available, this report will be supplemented.
 
Manufacturer Narrative
This supplemental report is to inform that upon further review, per the legal manufacturer, there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.Initially, we determined that the event was mdr reportable due to a backflow of liquid from the water container into the unit co2 regulator (ucr), in which case we determined that it was a potential adverse event because of the risk for infection.Upon further investigation, it was found that liquid does not flow back into the ucr from the water container unless multiple situations occur simultaneously.There are no reports of situations occurring for this complaint.In addition, a component analysis was performed on the water droplet traces in the tube of the ucr at a similar complaint.The results of this component analysis detected silica and others that appeared to be derived from tap water, but no component that appeared to be body fluids were detected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPIC CO2 REGULATION UNIT
Type of Device
CO2 REGULATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11337947
MDR Text Key232231988
Report Number8010047-2021-02759
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K081173
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberUCR
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/28/2021
Initial Date FDA Received02/17/2021
Supplement Dates Manufacturer Received03/08/2021
06/06/2022
Supplement Dates FDA Received03/31/2021
06/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-