• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported the vitrectomy stopped during surgery.There was no system message displayed.There was no patient impact.Additional information has been requested but not received.
 
Manufacturer Narrative
Additional information has been provided in d.10., h.3., h.6.And h.10.The company service representative examined the system and replicated the reported event.The nonconforming pneumatics module was replaced to resolve the issue.The system was then tested and met all product specifications.The system manufacturing device history record (dhr) was reviewed.Based on assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to the nonconforming pneumatics module.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key11338402
MDR Text Key233732015
Report Number2028159-2021-00155
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 01/27/2021
Initial Date FDA Received02/17/2021
Supplement Dates Manufacturer Received04/16/2021
Supplement Dates FDA Received05/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-