Model Number 1365-51-000 |
Device Problems
Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Ossification (1428); Abscess (1690); Adhesion(s) (1695); Fever (1858); Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Unspecified Infection (1930); Inflammation (1932); Necrosis (1971); Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Distress (2329); Osteolysis (2377); Thrombosis/Thrombus (4440); Joint Laxity (4526); Metal Related Pathology (4530); Unspecified Tissue Injury (4559); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 09/24/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Ad (b)(6) 2021: unfiled claim and the medical record received.After review of the medical records, the patient was revised to address septic right hip with suspected osteomyelitis of the innominate and the femur and allergic reaction to metal debris with significant muscle and skeletal necrosis.Operative note reported bone loss, groin pain.And infection.It was also mentioned that there was greenish to yellow thin fluid present.The trunnion was significant for black debris that was present even inside the head that most likely represented metallosis.It was also noted that there was advanced trunnionosis.Doi: (b)(6) 2006, dor: (b)(6) 2019, right hip (first revision).Please see (b)(4) for the third revision.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h6 (medical device problem code).
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : a device manufacturing (mre) review will not be performed even when product/lot information is known.Per wi-3430 it has been determined that, for the mom platform and related allegations an mre is not required.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H6 health effect - clinical code: appropriate term / code not available (e2402) used to capture the metal poisoning and blood heavy metal increased.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Litigation alleges that patient experienced elevated metal ion leaving a ¿very large dead space to manage.Plaintiff has sustained and will continue to sustain severe and debilitating injuries, economic loss, and other damages including, but not limited to, cost of medical care, rehabilitation, permanent instability and loss of balance, immobility, and pain and suffering.Plaintiff has sustained and will continue to sustain severe physical injuries, severe emotional distress, mental anguish, physical disability, economic losses, and other damages, as set forth in this complaint.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : a device manufacturing (mre) review will not be performed even when product/lot information is known.Per wi-3430 it has been determined that, for the mom platform and related allegations an mre is not required.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Medical records were received and stated the following: on (b)(6)2019, a progress reported that patient presented with worsening pain and developed fever.Patient was diagnosed infection and inflammatory response of the internal right hip prosthesis.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned to depuy synthes for evaluation.Review of the x-ray evidence was not able to confirmed the reported allegation.In order to observe implant wear, a photo of the explanted device is required.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
In addition to what was previously reported in the medical records the patient was revised to address osteolysis.Hip pain, increased polys, bone loss, nonviable abductor, adductor muscle, abscess, scar tissues and hip infection.Operative note reported significant bone loss in the innominate and the femur.The patient had a marked amount of muscle necrosis extended inside the pelvis for at least 10 cm, involved the iliacus and the psoas.The patient also had involvement of the abductors and adductors.Bone loss about the innominate was significant from ischium to supratectal region with significant anteromedial defect and posterolateral defect.The patient had a previous hardinge approach and there was a significant damage to the abductors and very minimal attachment present.Patient had dvt and underwent filter to decrease the risk of pulmonary embolus but still will require anticoagulant treatment and will have a significant risk of bleeding into the pelvis.Muscle loss and the dead space would be certainly difficult to manage.There was limited amount of adductor musculature and most of this was scar present.There was greenish to yellow rather thin fluid present and under pressure about 1 to 2 quarts of fluid was aspirated.This shrank the size of the pelvis considerably.Removal of the trunnion had black debris that was present in the femoral head, around the trunnion, tissue and at the neck.Most likely presented metallosis at the cobalt chrome trunnion interface.Clinical visit reported heterotopic ossification.Debridement of intrapelvic.Psoas and iliacus were completely necrotic, muscle was completely nonviable and avascular.Additional debridement performed at the adductor interval at the patient's previous site of surgery.There as abscess sinus formation in this area over the troch.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : according to the information received, "ad (b)(6) 2021: unfiled claim and the medical record received.After review of the medical records, the patient was revised to address septic right hip with suspected osteomyelitis of the innominate and the femur and allergic reaction to metal debris with significant muscle and skeletal necrosis.Operative note reported bone loss, groin pain.And infection.It was also mentioned that there was greenish to yellow thin fluid present.The trunnion was significant for black debris that was present even inside the head that most likely represented metallosis.It was also noted that there was advanced trunnionosis".The product was not returned to depuy synthes, however photos were provided for review.(b)(4).Review of the x-ray evidence was not able to confirmed the reported allegation.In order to observe implant wear, a photo of the explanted device is required.The overall complaint was unconfirmed as the observed condition of the articuleze m head 36mm +1.5 would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|
Search Alerts/Recalls
|
|