Model Number UCR
Device Problems Gas/Air Leak (2946); Pressure Problem (3012)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation.During the evaluation, the reported issue was confirmed due to a loose connection of gas tube.The output pressure was found out of specification due to a bad decompression unit.No other issues were found during the device inspection.If additional information is obtained a supplemental report will be filed.
 
Event Description
A user facility reported a leak causing and endoscopic co2 regulatory unit to lose co2 gas quickly.There was no patient involvement or injuries reported.No additional information has been obtained.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record (dhr) found no deviations that could have caused or contributed to the reported issue.The device met all specifications at the time of shipment.Based on the results of the investigation, although a root cause of the decompressor unit failure could not be conclusively determined, the probable cause is likely due to an accidental failure of the internal decompressor unit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPIC CO2 REGULATION UNIT
Type of Device
CO2 REGULATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11339381
MDR Text Key232799805
Report Number8010047-2021-02769
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
PMA/PMN Number
K081173
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 03/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUCR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/19/2021
Initial Date FDA Received02/17/2021
Supplement Dates Manufacturer Received03/09/2021
Supplement Dates FDA Received03/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-