• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problems Device Alarm System (1012); Device Displays Incorrect Message (2591)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2021.Date of this report: 17feb2021.
 
Event Description
It was reported that while the ventilator was in operation "backup alarm failed" was displayed and an alarm sounded.The ventilator was in use with a patient at the time of the alarm, and the patient was swapped to a different ventilator.No additional patient harm was reported.
 
Manufacturer Narrative
G4:08mar2021.B4:13mar2021.The field service engineer (fse) performed periodic maintenance.The reported issue was unable to be confirmed by an operational check performed.Since the backup alarm failed error was confirmed in the event log, the central processing unit board that had a backup alarm function was replaced by the fse to prevent a recurrence.The issue is resolved.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Manufacturer Narrative
The central processing unit printed circuit board assembly was returned for failure investigation.The piezo buzzer (ls1) was failing intermittently due to the insulation resistance was out of specification 457 kohms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key11339483
MDR Text Key233796627
Report Number2031642-2021-00629
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup,Followup
Report Date 08/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/21/2021
Initial Date FDA Received02/17/2021
Supplement Dates Manufacturer Received01/21/2021
07/19/2021
Supplement Dates FDA Received03/13/2021
08/16/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/26/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN PATIENT CIRCUIT, MASK, AND HUMIDIFIER
-
-