Additional information was added to d9, h3, h4, and h6.H4: device manufactured between june 26, 2020 to june 29, 2020.H10: the device was received for evaluation.A visual inspection was performed.And it was noted, that there was a tear at the lower left side edge of the bag.Functional testing was performed.And it was noted, that there was a leak at the lower left edge of the bag, just above the left side shoulder port weld.A magnified visual inspection was also performed.And the tear/hole in the lower left edge of the bag was verified.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted, and there were no deviations found related to this reported condition, during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|