Model Number V60
Device Problem Key or Button Unresponsive/not Working (4063)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2021.Date of this report: 17feb2021.
 
Event Description
The customer reported a ventilator was identified with a defective nav-ring.The device was not in clinical use at the time the issue was discovered.There was no patient or user harm reported.
 
Manufacturer Narrative
G4:05mar2021.B4:13mar2021.The device was evaluated by the customer and philips field service engineer (fse) who confirmed the reported issue.The fse replaced the front bezel.Following the modification, the device was reported to fully meet specification and returned to use.No other anomalies were reported.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Manufacturer Narrative
The exact context of when the issue was found is unknown.Additional information regarding the context was requested, however, no response was received from the contacts.It was reported that the device was not in clinical use at the time the issue was discovered.There was no patient or user harm reported.A front bezel was returned for analysis.Previous investigations identified the cause of the navigation ring failure is due to liquid ingress due to variability of the assembly process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RICA
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
yobana sanchez
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key11340422
MDR Text Key233797196
Report Number2031642-2021-00640
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/22/2021
Initial Date FDA Received02/17/2021
Supplement Dates Manufacturer Received01/22/2021
11/10/2021
Supplement Dates FDA Received03/13/2021
12/03/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/19/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-