Model Number OER-4
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/28/2021
Event Type  malfunction  
Manufacturer Narrative
The device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined, because the device has not been returned to olympus medical systems corp.(omsc).However, there is possibility that the reported event was caused by the following; an object was fallen on the lid.The effect of repeating pushing more than necessary when closing the lid.Mechanical stress when closing the lid and stress corrosion cracking (solvent cracking) caused by not wiping off chemical substances (detergents, chemicals, etc.) dropped on the lid.Some other factor.Device history record (dhr) review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed by the user that the lid of the device was cracked.There was no report of patient injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OER-4 100V
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11341304
MDR Text Key232592141
Report Number8010047-2021-02801
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-4
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/28/2021
Initial Date FDA Received02/17/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/29/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-