• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HUDSON CONCHA NEPTUNE; HUMIDIFIER, RESPIRATORY GAS, (

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HUDSON CONCHA NEPTUNE; HUMIDIFIER, RESPIRATORY GAS, ( Back to Search Results
Model Number IPN913776
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/26/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event Description
It was reported that the unit "is shutting down".No patient involvement reported.
 
Event Description
It was reported that the unit "is shutting down".No patient involvement reported.
 
Manufacturer Narrative
Qn#(b)(4).The sample was returned for evaluation.A visual exam was performed and there were no defects observed.Functional testing was performed and the unit was connected to 110vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.The unit was prepared for a full functional bench test in attempt to replicate the reported defect.A water reservoir bottle, an adult breathing circuit (880-36kit), a 382-10 conchasmart column, and dual temperature probes were connected to the unit.2-3 lpm of air flow were supplied to the unit for a real time operational scenario.The settings were set to full rainout, invasive, at 37 degrees c.The unit shut off during the test.The unit was opened and the power supply board was replaced with a known good lab inventory power supply board.This time, the unit successfully navigated all pre-operational self-tests again and was functioning in real time.The unit functioned without any interruption and was able to heat up to the set temperature of 37 degrees c with the lab inventory power supply board.Testing confirms the unit had a faulty power supply board.The complaint has been confirmed.The investigation revealed a defective power supply board.The root cause for the failure is unknown.The power supply board may have encountered higher stresses than normal causing it to fail before its intended use life.Teleflex will continue to monitor and trend on complaints of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHA NEPTUNE
Type of Device
HUMIDIFIER, RESPIRATORY GAS, (
MDR Report Key11342775
MDR Text Key232380292
Report Number3003898360-2021-00129
Device Sequence Number1
Product Code BTT
UDI-Device Identifier14026704646811
UDI-Public14026704646811
Combination Product (y/n)N
PMA/PMN Number
K131912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 01/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberIPN913776
Device Catalogue Number425-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2021
Initial Date Manufacturer Received 01/28/2021
Initial Date FDA Received02/18/2021
Supplement Dates Manufacturer Received03/18/2021
Supplement Dates FDA Received03/19/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.
-
-