COOK IRELAND LTD EVOLUTION® COLONIC CONTROLLED-RELEASE STENT - UNCOVERED; MQR STENT, COLONIC METALLIC EXPANDABLE Back to Search Results
Catalog Number UNKNOWN
Device Problems Obstruction of Flow (2423); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Occlusion (1984); Obstruction/Occlusion (2422)
Event Date 02/22/2020
Event Type  Injury  
Manufacturer Narrative
Pma/510(k) #: k163468.Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
 
Event Description
Lovero-2020, evo-c, endoscopic stenting for colorectal cancer obstruction as a bridge to surgery strategy.In an 8-year period, consecutive patients with acute left-sided colonic obstruction, due to locally advanced colorectal cancer, underwent sems implantation.After the stenting, all patients underwent surgery as an elective procedure.The type of surgery, was left at the surgeon's preference.Of the 22 stents implanted, 2 were evolution colonic stents.Complications in the patients receiving sems were: stent occlusion, perforations, migration, obstruction due to cancer ingrowth.In 2 patients, occlusive symptoms persisted despite correct positioning of the stent at fluoroscopy.
 
Manufacturer Narrative
Pma/510(k) #: k163468.Device evaluation: the colonic devices involved in this complaint was not returned to cirl for evaluation.With the information provided, a document based investigation will be complete.This file was created from the attached journal article.Reference "lovero-2020 " complaint files (b)(4), (b)(4), (b)(4), and (b)(4) were opened as a result of this paper.This file (b)(4) was opened to investigate occlusion.Documents review including ifu review: as the rpn and lot number of the complaint stents are unknown, a review of the relevant manufacturing records cannot be conducted.However, prior to distribution evo colonic devices are subjected to a visual inspection and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.As the evo colonic devices are from an unknown lot number, a review of the relevant manufacturing records cannot be conducted.As per the instructions for use, ifu (b)(4), which accompanies this device it informs the user about the potential complications "additional complications include, but are not limited to: tumor ingrowth or overgrowth , stent occlusion ¿ there is no evidence to suggest that the customer did not follow the instructions for use.Root cause review: a definitive root cause could not be determined from the available information.A possible root cause could be attributed the patient¿s pre-existing condition.From the information provided it is known that the patient had locally advanced colorectal cancer.Summary: complaint is confirmed based on the customers testimony.According to the initial reporter, patient successfully underwent follow-up surgery.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event Description
Supplemental report is being submitted due to the completion of the investigation.
 
Manufacturer Narrative
Pma/510(k) #: k163468.Device evaluation: the colonic devices involved in this complaint was not returned to cirl for evaluation.With the information provided, a document based investigation will be complete.This file was created from the journal article."lovero-2020 ".Complaint files pr (b)(4) (3001845648-2021-00129), pr 320039 (3001845648-2021-00123), pr (b)(4) (3001845648-2021-00127), and pr (b)(4) (3001845648-2021-00135) were opened as a result of this paper.This file (b)(4) (3001845648-2021-00129) was opened to investigate occlusion.Documents review including ifu review: as the rpn and lot number of the complaint stents are unknown, a review of the relevant manufacturing records cannot be conducted.However, prior to distribution evo colonic devices are subjected to a visual inspection and functional checks to ensure device integrity.As the evo colonic devices are from an unknown lot number, a review of the relevant manufacturing records cannot be conducted.As per the instructions for use, ifu0052, which accompanies this device it informs the user about the potential complications "additional complications include, but are not limited to : tumor ingrowth or overgrowth , stent occlusion ¿ there is no evidence to suggest that the customer did not follow the instructions for use.Root cause review: a definitive root cause could not be determined from the available information.A possible root cause could be attributed the patient¿s conditions occurring during device use.As per instructions for use, stent occlusion is listed as a complication following the placement of this device.However, as the device was not returned for evaluation; the cause of this complaint could not be conclusively determined.Summary: complaint is confirmed based on the customers testimony.As per medical advisor, patient outcome "require intervention/additional procedures" complaints of this nature will continue to be monitored for potential emerging trends.
 
Event Description
Supplemental report is being submitted due to a corrections to the completed investigation and multiple fields including imdrf coding.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUTION® COLONIC CONTROLLED-RELEASE STENT - UNCOVERED
Type of Device
MQR STENT, COLONIC METALLIC EXPANDABLE
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
Manufacturer (Section G)
COOK IRELAND LTD
o halloran road
national technology park
limerick
Manufacturer Contact
sinead o'leary
o halloran road
national technology park
limerick 
MDR Report Key11343007
MDR Text Key241431695
Report Number3001845648-2021-00129
Device Sequence Number1
Product Code MQR
Combination Product (y/n)N
Reporter Country CodeIT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Literature,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date02/22/2020
Event Location Hospital
Initial Date Manufacturer Received 01/22/2021
Initial Date FDA Received02/18/2021
Supplement Dates Manufacturer Received01/22/2021
01/22/2021
Supplement Dates FDA Received06/28/2021
09/24/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age71 YR
Patient SexMale
-
-